Download introduction - Venture Respiratory Inc.

Transcript
EN
specifications
Filter Specifications
Standard Filter > 3.0 micron particles
Optional Fine Particle Filter > 0.3 micron particles
Maximum Flowrate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH20
Measured
pressure at the
patient connection
port (cmH20)
Average flow at
the patient
connection port
(l/min)
7.0 cmH20 12.0 cmH20 16.0 cmH20
20.0 cmH20
2.9
6.4
11.0
15.0
19.0
83.8
126.8
137.4
130.0
113.8
Long Term Static Pressure Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . +/- 0.5 cmH2O
Dynamic Short-Term Pressure Accuracy with and without
DV5HH Humidifier (per ISO 17510-1:2007)
Waveform
Sinusoidal cycle with
I:E = 1:1
Volume
[mL]
500
Breath Rate
[min-1]
Set Pressure
[cmH2O]
Pressure Accuracy
[Pk-Pk cmH2O]
10, 15, 20
3.0
7.0
12.0
16.0
20.0
0.5
0.5
1.0
1.0
1.0
Equipment classification with respect to protection from electric shock . . . . . . . . . . .Class II
Degree of protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Part
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX1
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air
or with oxygen or nitrous oxide.
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
30
A-DV53