Download KRYPTON Help Pages on DMM Modular

Transcript
page 30 / 98
Click “OK”, each LED is listed as a blue frame in the overview. Each LED has a default
name, being its frame name extended with the LED sequence number in the frame
definition.
You can change this default name by double-clicking on the LED name and overtyping it
with a new name.
All chosen dynamic frames are listed in the big text field in the middle of the screen.
To remove a frame, select the frame until it is high-lighted and click “Remove 6D frame”.
The frame will be removed from the text field. Its LED’s remain configured, albeit as
single LED’s !
If frames are configured, the software will automatically calculate the frame motion from
the motions of the frame’s single LED’s.
9.3.
Tab ‘Analog’
This tab is only relevant if you have an ADC600 for measuring analog inputs.
If you have no ADC600, proceed with chapter “Measuring single LED’s or objects” on
page 32.
To enable or disable all 16 channels at once, check/uncheck the “Enable” check-box on
top.
Configure for each channel the following:
Figure 9: measurement configuration - tab 'analog'
Enable: enable/disable the individual channel
Name: keep the default name, or specify your own channel name by double-clicking and
overtyping the name.
Unit: specify the unit you want the result to be expressed in
Gain: select the amplification factor of the channel. Use only one of the factors you can
select from the combo box.
Bipolar: uncheck to only allow unipolar signals ( but with a doubled resolution ).
Remember that unipolar signals need a gain of at least x2 !
Differential: not implemented yet
You can have the result immediately as a linear function of the voltage: Y = M * X + Q
Factor: “Factor” is the “M”, the linear scale factor. This scale factor should NOT be
altered when changing the gain ! The gain is applied only to the analog signal. Once
the analog value is digitized, the numerical value is divided again by the same gain
factor. Therefore, the scale factor is always applied to a value as if the gain was X1.
Offset: “Offset” is the “Q” in the linear scaling
KRYPTON Help Pages on DMM Modular