12 • The Qualification Programs
Creating a Gradient Profile Data File Using a UV Detector
Preparing the System
Notes
• Use only ASTM Type 1, 18 MΩ or better, filtered, deionized
water to prepare eluents
• Eluents must be degassed and pressurized with helium.
1. Connect the gradient pump transducer outlet tubing to the
absorbance cell inlet port using about 25 cm (10 in) of yellow
0.076-mm ID (0.003-in) backpressure tubing (P/N 049715).
Backpressure must be over 7 MPa (1000 psi).
2. Connect the absorbance cell outlet to waste using a 75 psi
backpressure regulator (PEEK regulator, P/N 046480; stainless
steel regulator, P/N 039670) or enough 0.25-mm (0.010-in)
backpressure tubing to create 0.5 MPa (75 psi).
3. Set the UV detector wavelength to 254 nm.
4. Prepare the following eluent: 0.3% v/v acetone / 99.7 %
deionized water.
5. Insert eluent lines A and C into one bottle containing the eluent
prepared in Step 4.
6. Insert eluent lines B and D into one bottle containing deionized
water.
7. Prime each of the four eluent lines. Then, select 100% eluent B
and a flow rate of 1 mL/min (for standard bore) or 0.75 mL/min
(for microbore).
8. Allow the system to stabilize for at least 10 min.