PSI System Controls and Diagnostics
where ADC cards or counter cards are already available. If you use these outputs exclusively for
taking data, then the communications interface is only needed for setting up the device.
By default, the four analog monitors put out an analog voltage that maps the full scale bipolar
current range that is being used. Similarly the frequency monitors put out a frequency that maps
the same full scale. Since we can’t generate a negative frequency, the I404 offers two options for
mapping the signal current range onto the frequency range, as illustrate in figure 29. Normal
mode maps the full scale current range onto the 0 to 1 MHz frequency range, so that the
frequency will be 500 kHz for zero current. Absolute mode puts out 0 Hz at zero current, and
maps increasing absolute current to frequencies up to 1 MHz. Clearly you must be able to
resolve the two-valued response by other means in this case.
+10 V
-FS
1 MHz
1 MHz
500 kHz
500 kHz
+FS
-10 V
-FS
0 Hz
+FS
-FS
0 Hz
+FS
Figure 29. Monitor output mapping - current signals. Analog voltage on the left, Normal and
Absolute frequency mappings on the right.
In firmware releases prior to 4.1O, Normal mode was called Bipolar mode, and there were
Positive and Negative modes which mapped only positive or negative currents onto the output
frequency range. It is still possible to emulate these legacy modes using the compensation
parameters.
You have considerable flexibility to change the monitor mappings of measured current using the
sensor gain and offset compensation parameters (see section 14). This allows you to compensate
for differences between sensors, for example. The factors only affect the monitor outputs if you
select this to happen (for example by checking the relevant box in the PSI Diagnostic program
), otherwise they only affect the values that feed into the
position calculations.
If you choose to let the compensation factors affect the monitor outputs, then the analog voltage,
V, normal mode frequency, FN, and absolute mode frequency, FA, are given by:
V = 10.G
( I meas + O )
FS
+ O) 
 (I
FN = 5e5. G. meas
+ 1
FS


I404 User Manual
I404_UM_130129
Page 51 of 104