4 Installation
4.1 Site preparation
4.1.3 Site environment
4.1.3
Site environment
Introduction
This section describes the environmental requirements for installation of ÄKTA pure.
Operating conditions
The following requirements must be fulfilled:
•
The instrument is intended for indoor use only.
•
The room must have exhaust ventilation.
•
The instrument should not be exposed to direct sunlight.
•
Dust in the atmosphere should be kept to a minimum.
Allowed operating conditions are specified in the table below.
Parameter
Allowed range
Ambient temperature, operating
4°C to 35°C
Ambient temperature, storage
-25°C to +60°C
Relative humidity, operating
20% to 95%, non condensing
Altitude
Maximum 2000 m
Pollution degree
2
Note:
Do not use buffers with a freezing point close to or below the temperature in
the room or in the cold room cabinet.
Heat output
The heat output data is listed in the table below.
Component
Heat output
ÄKTA pure instrument
Typically 300 W
Maximum 600 W
Computer, incl. monitor and printer
Typically 300 W
Refer to manufacturer's instructions for
more information.