RS-232 Communications Interface
Note: The controller may still “talk” to the Model 512 during this time since a query
is not involved. The controller would detect 0 PC (Operation Complete) by
sending ENQ from time to time (ESB is enabled). When the ESB-bit goes
TRUE on EOI, RQS is set in the Status Byte, and the controller can then start
sending the necessary commands to dump data.
*OPC? The *OPC? command tells the Model 512 to place a ‘1’ into the Output
Queue (it will be transmitted if XON is in effect) when it reaches End Of Interval; the
MAV bit in SRE is then set (if enabled). The controller can detect this by ENQuiring
the Status Byte.
Note: Since this is a query, the controller cannot talk to the Model 512 while waiting
for End Of Interval. Until the Model 512 has responded, sending anything will
cause a Query Error.
*OPC? should be the last (and preferably the only) query in the Program Message, because the <RMT> is not sent until the ‘1’ is placed into the Output Queue.
EOI Bit The EOI bit in the status byte is set when the Model 512 reaches End Of Interval, if the EOI bit is enabled in SRE. The controller detects this by ENQuiring the
Status Byte from time to time. When EOI goes TRUE, the controller service routine
could query the Model 512 for events, time, counts information.
Note: Since a query is NOT involved, the controller can still talk to the Model 512
while waiting for End Of Interval.
Mode+Option Program Option
The Program Option function is enabled via the MODE+OPTION sequence (the
Model 512 must be in LOCAL mode). When MODE+OPTION is first pressed, the
OPTION push-button illuminates, and the message:
PROGRAM
OPTION
appears for two seconds, after which, the message:
SetUp232
appears in the upper display; the first selection of the option menu, “BaudRate”,
appears in the lower display.
SetUp232 Menu
(Upper Display)
54