Download POU 3300 with PSC - LAUDA-Noah

Transcript
Noah Precision, LLC
DANGER:
RISK OF DEATH OR BODILY INJURY. Disconnect all sources of input power
before working on this unit or anything connected to it.
Conditions of Use
To be in compliance with the stated directives and standards, the following conditions of
use must be met.
•
•
•
•
•
•
1-4
Operate the Model 3300/PSC HQ only with the type of power connector (190-230
VAC, 30 A, 47-63Hz, 4-wire connection)
Install and operate this device in an over voltage category II installation only.
Install and operate this device only in a pollution degree 2 or better environment,
which means an indoor location such as a computer room, office, or factory floor
where only non-conductive pollution occurs during operation. Occasionally, a
temporary conductivity caused by condensation occurs when the device is not
operating.
To prevent toppling, bolt the rack-mounted power supply/controllers into the rack
and bolt the rack to a floor or wall. The 3300 module must also be securely mounted
by a bracket or in a subfloor tile mount to prevent toppling.
Install the Power Distribution Box so that the input power connection is inaccessible
to the user behind the electrical panel.
Use only shielded cables for RS-485 communications.
Introduction