Hukseflux Thermal Sensors
A thermopile essentially is a number of thermocouples in series.
A thermocouple delivers an output signal that is proportional to
the differential temperature between the hot joints and the cold
joints. Multiple thermocouples in series, a thermopile, will
produce a proportionally larger signal. In case of TP01 the hot
joints are located near the heating wire (at 1 mm distance, rh )
and the cold joints are located far away from the heater (at
about 5 mm, rc). There are two rows of each 20 thermocouples
(copper – constantan) , which results in a signal of about 1,6 mV
when the medium at 1 mm from the heater differs 1 degree
Celsius from the medium at 5 mm from the heater.
This sensitivity is not equalled by any other sensor that is known
to us. It opens the possibility to reduce the sensor dimensions
considerably and to use low heater power, which is essential for
accurate measurements in humid soils. In humid soils (so not
saturated and also not perfectly dry) it is recommended that the
heater power remains low to avoid local transport of moisture by
evaporation. The 0.8 W/m of TP01 is the recommended
maximum.