Download Sigma Delta User Manual
Transcript
INSTALLATION 6.7 Pre-Use Check List In addition to the pre-use warnings listed for different models in sections 6.2 to 6.6 the following check list procedure must be carried out on ALL vaporizers before use. 1. 2. 3. 4. 5. Check that the vaporizer concentration control is in the 0 (zero) position. Check that the liquid level is between the upper and lower marks on the filler block. On key fill (agent specific filler) models, check that the filler plug is fully inserted and that the clamp screw is fully tightened. On pour fill (screw cap filler) models, check that the filler cap is securely closed. Perform a back bar manifold leak test as detailed in the relevant anaesthetic machine user instruction manual. WARNING Anaesthetic machine designs are constantly evolving and new models may differ dimensionally from existing equipment. It is the user's responsibility to ensure that the configuration of the anaesthetic machine allows correct installation of the vaporizer. 27