Transcript
Setup Process Reference Enter the conformity test file name either manually into the File name entry field, or click on the Load button to select the conformity test file from the respective directory. Conformity test files have the extension *.CFT. Define the path of the conformity test method. If you have selected the conformity test method file name by means of the Load button, the path is automatically entered into the Path entry field. In principle, the conformity test method may be stored in any path but it is recommended to either use the OPUS directory or a well-defined sub-directory of the process scenario path. 6.10.1 Result There are different possibilities for the channel result output: • Flag if conformity test passed If you activate this option button, the channel result will be identified, i.e. 1 means test has passed, 0 means test has not passed. • Max Cl value If you activate this option button, the maximum conformity index value will be displayed as channel result. • Sum 1 If you activate this option button, the result will be displayed on the basis of CI > limit (/N total). • Sum 2 If you activate this option button, the result will be displayed on the basis of CI > limit (/N over limit). For detailed information on the conformity test, refer to the OPUS IDENT manual. 60 OPUS/PROCESS Bruker Optik GmbH