Download Aurora 1000
Transcript
Maintenance Aurora 1000 User manual 1.3 Figure 48 Measurement chamber right cylinder components 15. When this has been completed replace chamber cylinder. 16. Remove the end cap of the left side measurement chamber cylinder (light trap) by pulling it off and away from centre (may need some twisting). Figure 49 Removing Light Trap 17. Clean light trap with a lint free tissue and warm water, leave to dry then replace back on the instrument. Clean the O-ring if necessary, however avoid using O-Ring grease. 18. Replace cell into instrument and replace all connections in reverse order to removal. 19. Perform a full calibration before normal operations can begin. 68