Technical Data
V-MON X represents the servo-control output X-Out, whereas V-MON Y
represents the servo-control output Y-Out. SGS-MON X represents X-In,
SGS-MON Y represents Y-In.
NOTE - RANGE OF CONTROL INPUT FOR
E-616.SS0 AND E-616.SS0G
By default E-616 Controllers are preset for closed-loop mode. There
are different ranges of control input for closed-loop and for open-loop
mode.
For E-616.SS0 and E-616.SS0G following control input ranges are
valid:
■
X- Input, Y-Input, closed-loop: 0 to +10 V
■
X-Input, Y-Input, open-loop: -2 to +12 V
If you apply voltage exceeding these ranges the overflow LEDs glow.
The amplifier output voltage ranges from -20 V to + 120 V.
www.pi.ws
E-616 PZ200E Release 1.0.2