Download MERCURY Acceptance Test Procedures and Specifications
Transcript
4.5 Magnet Drift 4.5 Magnet Drift Table 17 lists the drift specifications for magnets. Specifications for nominal field decay rate are less than or equal to the values listed in the table. About Magnet Drift Specifications Use 1H lineshape sample 20% CHCl3 in 80% deuteroacetone (CD3)2CO, Part No. 00968120-76. We recommend using a 5-mm probe capable of 1H direct observe. Table 17. Magnet Drift Specifications System (MHz/mm) Field Strength (T) Nominal Field Decay Rate (Hz/hr) 200/54, 200/89 4.70 2 300/54, 300/89 7.05 3 400/54 9.40 8 400/89 9.40 10 87-192327-00 E0997 MERCURY Acceptance Test Procedures and Specifications 71