VivoSense – Complex Respiration Analysis Module
Respiratory Drive Measures:
Vt/Ti: Mean Inspiratory Flow
The mean inspiratory flow is a measure of respiratory center drive. The higher its value, the greater the drive
and vice-versa. In situations in which thoracoabdominal discoordination is present, Vt/Ti may underestimate
drive since it does not take into account the drive needed to move the paradoxically moving compartment. In
obstructive apneas where respiratory efforts take place, the value of Vt/Ti approaches zero and therefore
falsely indicates that no drive is present. In this situation, respiratory drive is better measured from drive
measures of the RC or AB compartments.
Respiration Trends
Trends exist for all the above respiration measures contained in mirrored trend folders.