Download NO - BIOSENSOR USER MANUAL
Transcript
OVERVIEW This manual covers the use of the biosensor for nitrate/nitrite (NOx-) and for nitrite (NO2-). For simplicity, this will be refered to as NOX- in this manual. The NOx- biosensor consists of an electrochemical component (the nitrous oxide transducer) and a biological component (the biochamber). In order to optimize the lifetime of the sensor the biochamber is replaceable. Replaceable biochamber Cable includes guard Nitrous oxide transducer IMPORTANT Read the technical description of the NOx- biosensor before starting LEMO plug Fig. 1. Nitrate biosensor with replaceable biochamber Nitrate or nitrite diffuses into the NOx- biosensor from the external environment through a selective membrane (Fig. 2). Bacteria situated in a reaction chamber behind this membrane reduce the nitrate or nitrite to nitrous oxide (N2O), which is detected by an electrochemical nitrous oxide transducer. The amount 9