Download TruSight Tumor Sample Preparation Guide - Support

Transcript
TruSight Tumor Protocol
Extension-Ligation of Bound Oligos
This process connects the hybridized upstream and downstream oligos. A DNA
polymerase extends the upstream oligo through the targeted region, and is ligated to the 5’
end of the downstream oligo using a DNA ligase. This results in the formation of products
containing the targeted regions of interest flanked by sequences required for amplification.
Estimated Time
} Total duration: 50 minutes
} Hands-on: 5 minutes
Consumables
Item
Quantity
Storage
Supplied By
ELM3 (Extension-Ligation Mix
3)
1 tube
-15° to -25°C
Illumina
Adhesive aluminum foil seal
1 seal
User
Troughs
As needed
User
Procedure
20
1
Using a multichannel pipette, add 45 µl of ELM3 to each sample well of the FPU plate.
The Extension-Ligation reaction takes place on the filter plate membrane.
Changing tips between columns is not required if you use care to avoid crosscontamination.
2
Seal the FPU plate with adhesive aluminum foil, and then cover with the lid to secure
the foil during incubation.
3
Incubate the entire FPU assembly in the pre-heated 37°C incubator for 45 minutes.
Ensure waste is removed from bottom of the plate.
4
While the FPU plate is incubating, prepare the IAP (Indexed Amplification Plate) as
described in the following section.
Part # 15042911 Rev. A