Download moisture analyzer ma 3y series
Transcript
Means of setting each parameter is the same as in case of the previous menu option. The only difference refers to possibility of declaring parameter of controlling start mass for a sample. This parameter determines use of a sample with specific mass to test its moisture content. CONTROL OF START MASS enables setting one of three options: No – option disabled, mass of a sample prepared for drying is not controlled by the software Optional – option enabled, the display shows a bar graph demonstrating graphic visualization of sample mass in relation to mass value saved in the database of products. It is a sample mass and its tolerance limits set in the software. Before applying this option, add a product in the database, including data on sample mass and its checkweighing limits. On enabling this option, the moisture analyzer’s software displays commands on placing a sample with determined mass on the drying pan. However, if a user decides otherwise, then a moisture analyzer starts drying process even if sample mass exceeds determined threshold limits for a tested sample mass. Mandatory - option enabled, the display shows a bar graph demonstrating graphic visualization of sample mass in relation to mass value saved in the database of products. It is a sample mass and its tolerance limits set in the software. Before applying this option, add a product in the database, including data on sample mass and its checkweighing limits. On enabling this option, the moisture analyzer’s software displays commands on placing a sample with determined mass on the drying pan. Unless the sample mass is within set threshold limits, the software does not allow to start drying process. There are two ways of using this option: 1. For < an optional product, using Checkweighing thresholds> or < quick access keys: Target value>. 2. In combination with a specific product saved in the < products>. Database of In the 1st case, the target value, checkweighing thresholds and percent tolerance of mass should be set using quick access keys. In the 2nd case, the value of mass, and tolerance fields have to be specified their corresponding fields of the <Database of products>. - 70 -