Means of setting each parameter is the same as in case of the previous
menu option. The only difference refers to possibility of declaring parameter
of controlling start mass for a sample.
This parameter determines use of a sample with specific mass to test its
moisture content.
CONTROL
OF
START
MASS
enables
setting
one
of
three
options:
No – option disabled, mass of a sample prepared for drying is not controlled
by the software
Optional – option enabled, the display shows a bar graph demonstrating
graphic visualization of sample mass in relation to mass value saved in the
database of products. It is a sample mass and its tolerance limits set in the
software.
Before applying this option, add a product in the database, including data on
sample mass and its checkweighing limits.
On enabling this option, the moisture analyzer’s software displays
commands on placing a sample with determined mass on the drying pan.
However, if a user decides otherwise, then a moisture analyzer starts drying
process even if sample mass exceeds determined threshold limits for a
tested sample mass.
Mandatory - option enabled, the display shows a bar graph demonstrating
graphic visualization of sample mass in relation to mass value saved in the
database of products. It is a sample mass and its tolerance limits set in the
software.
Before applying this option, add a product in the database, including data on
sample mass and its checkweighing limits.
On enabling this option, the moisture analyzer’s software displays
commands on placing a sample with determined mass on the drying pan.
Unless the sample mass is within set threshold limits, the software does not
allow to start drying process.
There are two ways of using this option:
1. For
<
an
optional
product,
using
Checkweighing thresholds> or <
quick
access
keys:
Target value>.
2. In combination with a specific product saved in the <
products>.
Database of
In the 1st case, the target value, checkweighing thresholds and percent
tolerance of mass should be set using quick access keys.
In the 2nd case, the value of mass, and tolerance fields have to be specified
their corresponding fields of the <Database of products>.
- 70 -