3 Entering Parameters Precooling Precooling Precooling is used to bring a rotor to a desired temperature prior to centrifugation of a sample. Precooling is defined as a set program and is called up via the configuration menu “Programs” on page 4-3. This program is activated using the button Load. In the main screen you can then set the required target temperature. The settings for acceleration and deceleration profile, time and speed are set from the centrifuge. If the values are changed, the precooling program is exited. Precooling: PRE-COOLING (marked line) is stored as a fixed program. Use the Load button to apply the program. You can set the target temperature after this in the main screen. 3-2 Touchscreen User Interface for Sorvall LYNX Thermo Scientific