Fig. 2-6
•
Press cartridge holder eject button (see Fig 1-7(6)) to eject the cartridge holder.
•
Place the provided Calibration Cartridge with barcode side facing up in the cartridge holder
and touch “START”.
•
mLabs® ImmunoMeter will perform calibration automatically. After calibration is completed,
mLabs® ImmunoMeter will display either “CALIBRATION PASS” or “CALIBRATION FAIL” on the
screen.
•
After calibration is complete, the calibration cartridge will be ejected. Remove the
calibration cartridge from the holder and replace it back to the original zip-locked pouch. DO NOT
DISPOSE OF THE CALBRATION CARTIDGE.
•
Press cartridge holder eject button to bring the cartridge holder back into mLabs®
ImmunoMeter.
•
Do not add sample or control or any liquid into the Calibration Cartridge!
•
If calibration fails, contact the local Micropoint distributor for further advice.