Download DL2e Data Logger Hardware Reference
Transcript
Appendices Appendix C: Calculating the speed of data readings NOTES: [1] SCAN OVERHEADS is the time taken for the logger to prepare for taking readings at the beginning of a LOG or EVENT, and performing ‘housekeeping’ calculations when finished. [2] The time required for an analogue reading depends on the magnitude of the input signal. ‘Full scale’ is with respect to the voltage range being used. If autoranging, the full scale figure applies for inputs of 4mV, 32mV, 262mV, 2.048V, and for corresponding resistance values (Vin = R x Iexcitation). [3] The timings quoted above apply for the 50/60 switch set to the European mains frequency, 50Hz. When set to 60Hz, all analogue timings are approximately 20% faster (see "Electrical mains environment" on page 48). [4] Autoranging to a less sensitive range is slower than to a more sensitive range, and requires an over-range recovery cycle in addition to the reading being taken. [5] The logger takes separate autozero readings for each terminal group, for voltage and resistance readings at each excitation current, as required for a LOG or EVENT. You can minimise autozero overheads for mixed resistance readings by grouping channels that use the same excitation current on consecutive input channels in the same terminal group. EXAMPLES: two autozero readings are required for channels 1 to 30 programmed as VLT channels: 5 autozero readings are required for channels 1 to 6 programmed as TM1, RR1, RR2, RR3, RR4, VLT, as in the DEFAULT configuration. [6] The logger takes a single cold junction reading for any number of thermocouple measurements in a single scan. DL2e Hardware Reference Page 151