5
CUSTOM CALIBRATION IN SPECIAL MODE
This process (made by the customer) allows to correct the “raw” signal provided by the sensor for a
given wire rope at a given strain. Contrary to a specific calibration made by the factory that gives a
correction with a formula.
A memory for 200 Custom calibrations is available.
In order to memorise the parameters of a given situation, prepare an installation as described below.
(Or insert a dynamometer in the installation on the actual site.) and follow the user manual instruction.
See : Calibration process in the Special Mode.
Calibration process
Fig 1
1.
fixed end
Tensioner
Working
load
Dynafor LLX
Wire rope
Identical to the actual installation
Length : from 0,5 m to ∞
Fixing end. flexible or rigid
according to the actual installation
Fig 2
Working load
+
load cell incidence
Bring the wire rope in contact
with the resting plate
100%
CAPTEUR
50%
H F 8 7 / P/ 1
10%
POWER
RS232
On/Off
S T are E
F1
F2
F3
Shift
Adjust and memorise the parameters
following the user manual.
See : Calibration process in Special Mode