If Dynamic Blank is not present and test is an End Point, then:
•
ABS = ABS – Blank Value
ABS = ABS x Fnr
•
•
If the factor is used, then: Conc = ABS x Factor
Otherwise the concentration is extrapolated from the standard’s curve, where:
Fnr : Internal Factor
ABS : Test ABS
Conc : Final concentration
1.2.2. APPLIED MATHEMATICAL FUNCTIONS
♦ Correlation Coefficient
∑ (T
n
CC =
i
1
n
∑ (L
i
− T )( Li − L)
− L)
2
1
n
∑ (T
i
− T )2
1
where:
n
i
T
L
: Number of readings
: Number of reading (i)
: Times
: Readings
♦ Linear Regression
n
n
∑ (T L ) − n
i
M=
n
2
i
1
n
i
1
n
∑ Li
1
n
Chapter C
n
1
1
∑ (T )
Section I
∑ (T )
i
n
Q=
n
∑ ( Li ) 2
2
−
(∑ Ti ) 2
1
n
n

 n
∑1 Li

− ∑ (Ti Li ) − n
 1
n


M
Functions
BT2000 PL

n
∑ T 
i
1
n 


Rev.0, Soft Ver.8