Download chapter 1 specifications
Transcript
Disassembly, Reassembly and Lubrication 3-6-11. “Unit, Control Panel”, “SA PNE Sensor”, “SA, Relay PCB” and “SA, Main PCB” (1) Unit, Control Panel 1. Remove the “Case U”. Refer to “3-6-3(1) Removing the “Case U” Block”. 3. Remove the “Unit, Control Panel” upwardly by releasing the hooks (2 pcs.). 4. Remove the “Cushion, Ope-Pane Cable” and disconnect the “Cable, Opepane” from the “Unit, Control Panel” (). Note on reassembling: • When connecting the flexible cable, be sure that the insertion direction is correct. If you insert it reversely, circuit is not electrically connected. (Its foil side faces to the right.) (2) SA PNE Sensor 1. Disconnect the connector (J12), remove the 1 screw (BH, M3x4 (NI)), and detach the “SA PNE Sensor” Block. 2. Remove the 2 screws (PH, M2x4) and detach the “SA PNE Sensor” () from the “Holder PNE”. BH, M3x4 (NI) BH, M3x4 (NI) SA PNE Sensor BH, M3x4 (NI) PH, M2x4 Spacer PNE BH, M3x4 (NI) Holder PNE PHT (#2), M3x6 J12 Cover Main PCB "Plate Main PCB" Block Cable, Opepane BH, M3x4 (NI) Cushion, Ope-Pane Cable Unit, Control Panel J4 J15 Wire Tie SA Power Cable 3-27 CL-S6621