Download PDA-100 (USB)
Transcript
3 • Operation and Maintenance 1200 1.02 mAU Bandwidth = 5 nm (270 nm to 274 nm) 1000 750 mAU 500 Reference Bandwidth = 51 nm (300 nm to 350 nm) 250 0 -200 220 nm 238 250 263 275 288 300 313 325 338 350 363 375 388 400 413 425 438 450 nm 272 nm Sample Wavelength 325 nm Reference Wavelength Figure 3-6. Caffeine Spectrum Bunch Width The bunch width setting determines how many nanometers are averaged when collecting 3D data. The default is 1 nm. Selecting a bunch width above 1 nm will reduce the required data storage by approximately the reciprocal. Do not select a bunch width when the spectrum of the compound has a fine structure; the bunch width will reduce the spectral resolution. Step Chromeleon automatically selects the appropriate step setting. Therefore, it is not necessary for users to select a step setting. A step is the time interval between two successively stored data points. The smaller the step, the more data points that are recorded and, in general, the more precise the analytical results. However, because collecting more data points requires more storage capacity, the step setting sometimes represents a compromise between the amount of information collected and the file size. Chromeleon automatically selects the step value that is the inverse of the data collection rate selected by the user. For example, if the data collection rate is 5.0 Hz, the software sets the step to 0.2 second. The following table indicates the step setting automatically selected for each data collection rate. Doc. 031898-03 3/06 49