Section 4 Operation System - Password/Reset is used only by a qualified technician. Changing the password enables circulator reset options, the temperature sensor calibration procedure and displays PID values. 1. With Password/Reset highlighted, press to display: 2. Press and change the number to 1. Level User Level User Password 0 Password 1 Menu Menu 3. Press to display: Note The circulator resets to the User mode when it is turned off. The circulator also resets to the User mode when the Start/Status Display is displayed continuously for 10 minutes. Level Operator Password 1 Reset Calibration Menu Scroll down to display PID Tuning. 1. If desired, highlight Reset and press to display: Reset user settings Reset PID settings Reset both Menu Highlight the desired reset option and press . A confirmation message appears, press again. 4-18 Thermo Scientific