Appendix A. CE Mark Compliance
Appendix A. CE Mark Compliance
A.1 Introduction
For CE Mark compliance, the Model GF868 flowmeter must be wired in accordance with the instructions in this
appendix.
IMPORTANT: CE Mark compliance is required only for units intended for use in EC countries.
A.2 Wiring
The Model GF868 must be wired with the recommended cable, and all connections must be properly shielded and
grounded. Refer to Table 10 for the specific requirements.
Table 10: Wiring Modifications
Connection
Transducer
Input/Output
Power
Shielding
Cable Type
Termination Modification
RG62 a/u
Add metallic cable clamp from braid to chassis gnd.
Armored RG62 a/u or conduit
None - grounded via cable gland.
22 AWG shielded (e.g. Baystate #78-1197)
Terminate shield to chassis ground.
Armored conduit
None - grounded via cable gland.
14 AWG, 3 conductor, shielded (e.g. Belden
#19364)
An external ground to the chassis is required.
Armored Conduit
None - grounded via cable gland.
For CE compliance, power and I/O cables must be shielded. Cables to be terminated within cable
gland at the GF868. Shielded cable is not required when installations include metal conduit.
A.3 External Grounding
For CE Mark compliance, the electronics enclosure and the transducer fixture must each have an external ground wire
attached.
Note: If the Model GF868 is wired as described in this appendix, the unit will comply with the EMC Directive
2004/108/EC.
DigitalFlow™ GF868 Startup Guide (1 and 2-Channel)
47