70032603 Revision 00
Source Wavelength
CH 1
OPM1: WAVLNGTH
1550.00 NM
NOTE:
Source wavelength must be entered in order for the
DPM-79810 to display the correct optical power.
Select Wavlngth using the PARAM key from the SETUP menu. Use the UP
or DOWN arrow key or the numeric keypad to enter the source wavelength.
Press the ENTER key. The range of available wavelength settings is
850nm to 1650nm, with a default value of 1550.00nm. If the source
wavelength is out of range the power reading will not be accurate.
Filtering Value (Averaging Rate)
CH 1
OPM1: FILTERING
1
The filter setting refers to the number of digital samples that are averaged
before a measurement is reported. This has the advantage of integrating
out the affects of noisy signals. Select Filtering using the PARAM key from
the SETUP menu. Use the UP or DOWN arrow key to enter the filtering
value. Press the ENTER key. The range of available filter settings is 1 to 50
samples (with 1 offering a measurement every 150msec and 50 offering a
measurement every 7.5 seconds).
Measurement Units
CH 1
OPM1: UNITS
MW
The optical power meter can display the measurements in either a linear
scale (Watts) or a Logarithmic scale (dBm). When viewing relative
measurements, either OPM1-OPM2 or OPM2-OPM1, the setting for
OPM1 will establish the units displayed. Select Units using the PARAM key
from the SETUP menu. Use the UP or DOWN arrow key to cycle between
the two available choices. Press the ENTER key. The available selection
for measurement units is either Linear (Watts) or Logarithmic (dBm).
5-12
CHAPTER 5 DPM-79810 MODULE: Operating Parameters Setup