When a dual detector is used with collection on two wavelengths, or when more than one m/zratios are specified when using mass detection, collection is started when one of the signals used
for collection and fractionation reaches the specified Start Threshold level. Fractionations occur
when another signal reaches the Start Threshold level and when one of the signals decreases below
the Start Threshold level. When all signals decrease below the Start Threshold level, the collection
is stopped.
Slope
The Slope collection method fractionates based on volume. When the collection is started depends
on the calculated slope that is based on the signal from the detector. The collection is stopped when
the signal is below the absorption level detected when the collection was started.
When a dual detector is used with collection on both wavelengths, collection is started when the
calculated slope of one of the signals used for collection and fractionation reaches the specified
Start Slope setting. Fractionations occur when the other signal’s calculated slope reaches the Start
Slope setting and then when one of the signals decreases below the absorption level detected when
the collection was started. When both signals decrease below the level detected when the collection
was started, the collection is stopped. The Slope collection method also enables the possibility to
fractionate when shoulders are detected in the light absorption curve.
It is possible to combine the Collect All with the Threshold and/or the Slope collection method. This
means that everything will be collected and fractionation will occur on the Start Threshold level
and/or the Start Slope setting. Fractionation will also occur on detected valleys as well as on
shoulders, if enabled.
Note that the Slope collection mode is only available with the UV1 and UV2 detection signals.
A.1.5
Collection and Fractionation Parameters
Method Parameters
The method parameters are defined for each method and can be changed by the user. The following
method parameters are available:
Table A-2: Method Parameters
Parameter
Description
Detection Signals
The basic configuration for collection and fractionation. The range is [λ-all | Mass |
UV1| UV2 | External].
The λ-all detection signal is only available on systems with an Isolera Spektra
license installed. The Mass detection signal is only available on Isolera Dalton
System. The UV2 detection signal is only available on systems with a dual
wavelength detector.
Start Wavelength
The shortest wavelength to be included in the λ-all signal. The range is 200 to 400
nm (dual variable UV detector) or 200 to 800 nm (UV-VIS detector). (Only
available with the λ-all detection signal.)
End Wavelength
The longest wavelength to be included in the λ-all signal. The range is 200 to 400
nm (dual variable UV detector) or 200 to 800 nm (UV-VIS detector). (Only
available with the λ-all detection signal.)
Ionization Mode
The ionization mode used for detecting mass ions can be positive or negative.
Mass 1-4
The mass-to-charge ratio(s) of the ion(s) used for detection, collection, and
fractionation. The range is 50 to 800 m/z.
Collect All
Used to collect the entire run. The range is [On | Off].
411829-R, Collection and Fractionation
April 2015