Download ABI 433A Peptide Synthesis
Transcript
Applied Biosystems Example of a Deprotection Calculation Module H (Piperidine Deprotection - Previous Peak) determines the number of deprotections based on the measured conductivity monitoring values. At the end of the third deprotection, the value of the second and third deprotections are compared to see if they are within the limits defined by step 54 (End loop monitoring). SynthAssist uses the following formula to calculate whether three or four deprotections are used: Value of 2nd deprotect – Value of 3rd deprotect = Value of 2nd deprotect percentage If the percentage calculated is lower than 2.5%, then no more deprotections will occur. For example, if the values of the three deprotections are 995, 775, and 766, 775-766 then = 775 9 775 = 0.012 = 1.2% And only three deprotections will occur. However, if the values of the three deprotections are 903, 792, and 771, 792-771 then 792 = 21 792 = 0.026 = 2.6% And four deprotections will occur in Module H. If four deprotections occur in Module H, then Module h will become active. Note November 2001 You can modify the value in step 54 to change the sensitivity of the monitoring. A lower value will provide greater sensitivity. 5 Cycles and Modules 5-11