Download ABI 433A Peptide Synthesis

Transcript
Applied Biosystems
Example of a Deprotection Calculation
Module H (Piperidine Deprotection - Previous Peak) determines the
number of deprotections based on the measured conductivity monitoring
values. At the end of the third deprotection, the value of the second and
third deprotections are compared to see if they are within the limits defined
by step 54 (End loop monitoring).
SynthAssist uses the following formula to calculate whether three or four
deprotections are used:
Value of 2nd deprotect – Value of 3rd deprotect
=
Value of 2nd deprotect
percentage
If the percentage calculated is lower than 2.5%, then no more
deprotections will occur.
For example, if the values of the three deprotections are 995, 775, and 766,
775-766
then
=
775
9
775
=
0.012
=
1.2%
And only three deprotections will occur.
However, if the values of the three deprotections are 903, 792, and 771,
792-771
then
792
=
21
792
=
0.026
=
2.6%
And four deprotections will occur in Module H.
If four deprotections occur in Module H, then Module h will become active.
Note
November 2001
You can modify the value in step 54 to change the sensitivity of the
monitoring. A lower value will provide greater sensitivity.
5 Cycles and Modules
5-11