Site Requirements and Specifications
Performance Specifications
Table 3
Performance Specification of the ELS Detector
Type
Specification
Communication
Ethernet
Serial (RS232)
Remote Start Input
Pump Stop: 1 Contact closure
PC control (software)
ELSD driver for OpenLAB ChemStation edition
ELSD driver for OpenLAB EZChrom edition
Remote operation
Remote Start Input
Safety features
Gas shut-off Valve, Leak Detection, Laser Interlock
Specification Conditions
ASTM: “Evaporative Light Scattering Detectors Used in Liquid
Chromatography”.
Table 4
Reference conditions
Gas flow
1.6 SLM
Neb temperature
40 °C
Evaporator temperature
40 °C
Data Rate
40 Hz
Smoothing
1
ASTM drift tests require a temperature change below 2 °C/h (3.6 °F/h)
over one hour period. Our published drift specification is based on these
conditions. Larger ambient temperature changes will result in larger drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency
and the amplitude of the temperature changes to below 1 °C/h (1.8 °F/h).
Turbulences around one minute or less can be ignored. ASTM
measurements require that the detector should be turned on enough time
before start of testing.