Download ML4000 User`s Guide - Ototronix Diagnostics

Transcript
Speech Discrimination
Name –The option to change the name of the test can be completed by typing an alternate test name in the name field under the
“Speech Discrimination” tab on the Otogram Tools screen.
Presentation Base Level – The Otogram will automatically determine the presentation level for this test by referencing the “Presentation Base Level”. The Presentation Base Level can be based on:
Articulation Index (AI) – Default setting used to maximize audibility of speech sounds.
Pure Tone Average (PTA) - words will be presented at PTA or minimum presentation level, whichever is greater.
Manual - the Otogram will prompt the administrator to enter the desired presentation and masking levels for each ear before beginning the speech discrimination test.
Boost – Should the presentation level need to be louder than the Presentation Base Level, it can be increased by the decibel
amount specified in the “Boost” field. Regardless of whether or not a boost value is specified, the words will never be presented
less than the Minimum Presentation Level.
Presentation Type – To test speech discrimination as an open-set test, change the “Presentation Type” from “Closed” to “Open”. If
the Otogram has been configured for open set speech discrimination, the administrator will be paged to return to the patient at the
appropriate point in the test battery to facilitate administration.
NOTE: This option is only available for English language tests; other languages must always use the automated speech discrimination test
Minimum Presentations – The minimum number of words presented by the Otogram for speech discrimination testing is 12. If the
patient gets the first 12 words correct, the test is complete. For each word missed, additional words are presented until a range of
statistically equivalent values (calculated at the 85% confidence interval) is reached.
Min Presentation Level—adjustable between 40 dB and 75 dB
Max Presentation Level—adjustable between 80 dB and 100 dB
Word List (Open Set, English Only)—option of NU-6 or W-22
Display Written Words—select this option to display written words beneath pictures
26