Download 9403-5 - Veterinary Instrumentation
Transcript
Go To Table Of Contents Back Next CARDELL® Veterinary Monitors 9. TEMPERATURE MONITORING YELLOW SPRINGS INTERNATIONAL (YSI) TECHNOLOGY The CARDELL Monitor is designed to measure temperature to aid in the assessment of thermoregulation. NOTE: For temperature monitoring a YSI temperature probe and Adapter are required. To take a Temperature measurement; 1) Consult proper procedure for application of the probe to the patient. 2) Connect the probe to the Temperature Adapter. Connect the Adapter to the TEMP connector on the side panel of the monitor. 3) Press the ON/STANDBY pushbutton to turn “ON” the monitor. 4) Check that the monitor is accurately measuring temperature. When the value has been determined, it will be displayed on the front panel as (oF or oC). 5) Configure the appropriate temperature scale (oF or oC). Refer to Page 95, SELECTING THE TEMPERATURE UNITS. NOTE: When the Temperature value being measured is greater than 43 oC (109.4 oF), the TEMP numeric value will be displayed as “>>>>>” indicating a value above the monitors temperature range. NOTE: When no Temperature probe is connected, the TEMP numeric display will be blank. Refer to Section 14, ACCESSORIES for Temperature probe types and part number information. Consult instructions enclosed with the sensor for proper application. 21-02-0186 REV. 08 07/11 75