Download HMG 3010 User Manual
Transcript
HMG 3010 Page 47 Linking Events You also have the option of linking events. Without linking, each of the four events acts alone and independently as a trigger for a measurement recording. In the line Joint events (options shown to the right) a drop-down list opens containing the linked event options. In the example, the linking of Events 1 and 2 (shown on the left by a circuit diagram symbol) means that recording is EITHER triggered by the simultaneous occurrence of events 1 and 2 (the measured value in channel A drops to 5.00 bar, with the measured value in channel B leaving the window of 4.00 to 6.00 bar at the same time) OR a voltage signal (falling signal edge) is detected at trigger input J, OR a key or button is actuated. Other event linking options are shown in the drop-down list. It should be borne in mind that events are always listed in descending order, otherwise you have to re-sort the events. In the 1 and 2 and 3 and 4 option, all events have to occur simultaneously in order for a recording to be triggered. By contrast the last option – 1 and 2 & 3 and 4 – provides three variants: A triggered recording is initiated when a) 1 and 2 coincide or b) 3 and 4 coincide or c) all, i.e. 1 and 2 and 3 and 4, occur simultaneously. Edition 2012-06-28 - V04 R01 Part No.: 669855