5.5 GO/NO-GO Determination I/O and External Start/Stop Input (EXT I/O)
Input Signal
START IN (Negative logic)
Use this signal to perform GO/NO-GO determination by synchronizing to an external input signal.
The signal is only valid when on the GO/NO-GO menu, Remote is set to ON. If Remote is set to
OFF, GO/NO-GO determination is performed regardless of the external signal input (the GO/ NOGO determination result is output).
Signal Input Circuit
5V
Equivalent to
LCX14
4.7 kΩ
START IN
120 Ω
HIGH
LOW
100 pF
START
Output Signal
NOGO OUT (Negative logic)
When the determination result is NO-GO (fail), the output signal level temporarily changes from high
(H) to low (L).
GO OUT (Negative logic)
When the determination result is GO, the output signal level temporarily changes from high (H) to
low (L).
Signal Output Circuit
5V
4.7 kΩ
100 pF
GO OUT
NO-GO OUT
120 Ω
GO/NO-GO I/O Timing
100 ms or more
50 ms
or more
START IN
Acquisition
Acquisition
Judgment
NOGO OUT
0 ms
or more
Result
Judgment
Result
GO OUT
START IN is not referenced when Remote
is set to OFF; GO OUT and NO-GO OUT
are output according to internal
GO/NO-GO determination.
5-6
The signal remains low until the DL850E/DL850EV
is ready to accept the next measurement.
The duration is longer during output to a built-in
printer or internal storage medium.