Primary User Interface Screens
Lactate Screen
On the Monitor screen, press the Lactate field (or the surrounding area) to bring
up the Lactate screen (see Figure 5-7). This screen allows an operator to view
the calculated lactate concentration in mg/mL, the percent CO2 in the offgas, and
the bicarbonate constant of the supply media. The bicarbonate constant is
calculated during the lactate calibration step of the PRE-INOCULATION
procedure, though an authorized user may manually change it. Biovest does not
recommend modifying the bicarbonate constant if the instrument is functioning
normally; changing this value may negatively affect the condition of the cell
culture environment.
Figure 5-7: Lactate Screen
5.1.9
Cycle Detail Screen
On the Monitor screen, press any cycling field (Cycle State, EC Pressure, or IC
Pressure) or the surrounding area to bring up the Cycle Detail screen (see Figure
5-8). This screen allows an operator to view the current cycling state, the userdefined cycling parameters (Cycle Delay Time, EC Rise Time, and EC Fall
Time), and the cycling pressures that will be utilized by the cycling algorithm
during the next cycle. An authorized user may change the Cycle Delay Time, EC
Rise Time, and EC Fall Time by pressing their respective fields. These
parameters are normally set on the Run Settings 1 screen (see Figure 5-19).