Download Sartocon® Slice 200 Cassettes
Transcript
Detection Method of NaOH cleaning agent clearance: −− Neutral pH effluent in both retentate and permeate. −− Conductivity. Example of a basic cleaning protocol for Hydrosart® Microfilters: Step 1: Flush the filter cassettes with isotonic saline for 5 min. at room temperature. Conditions: pin = 36.3 psi|2.5 bar|250 kPa pret = 0 psi|0 bar|0 kPa pper = closed; open after 4 min. Step 2: Recirculate 1 NaOH at 50°C for 10 min. Conditions: pin = 29 psi|2.0 bar|200 kPa pret = 0 psi|0 bar|0 kPa pper = open Step 3: Recirculate 1 N NaOH at room temperature for 10 min; then flush the filter cassettes with the same batch of sodium hydroxide. Conditions: pin = 29 psi|2.0 bar|200 kPa pret = 0 psi|0 bar|0 kPa pper = closed; open after 4 min. Step 4: Flush with purified water until the pH is neutral. Conditions: pin = 29 psi|2.0 bar|200 kPa pret = 14.5 psi|1.0 bar|100 kPa pper = 7.3 psi|0.5 bar|50 kPa Start up of cassettes 37