Introduction A standard of isobutylene at a known concentration may be used for setting the sensitivity. If the Multi-PID 2 is calibrated with isobutylene, it displays concentrations in units equivalent to ppm of isobutylene. If isobutylene were the only photoionizable chemical in the sample, then the Multi-PID 2 would display its concentration directly. The Multi-PID 2 responds more or less readily to other chemicals than it does to isobutylene. Because it has a medium sensitivity to isobutylene, this gas has been chosen as a reliable means of reporting an average concentration of total ionizable compounds present. Gases other than isobutylene can be used to calibrate the MultiPID 2. However, all response factors are based on an isobutylene calibration. Photoionization Detector The photoionization detector (PID) is shown in Figure 3. The PID measures the concentration of photoionizable chemicals in the gas stream from the sample inlet and produces an electrical signal for the microprocessor. A UV lamp generates photons which ionize specific molecules in the gas stream. The permanent air gases (argon, carbon dioxide, nitrogen, oxygen, water vapor, etc.) require a relatively high energy for ionization, and are not ionized by the UV photons. Many of the compounds considered pollutants, including most hydrocarbons, are ionized. The gas stream is directed into the PID through a small port at the center of the UV lamp window and through a series of larger ports around the perimeter of the lamp window. This arrangement permits a high sample flow rate and short response time. 21