6200 User’s Manual Pitot Leak mode and stop ‘Leak’ from blinking. Pressing CANC will revert back to Pitot Measure mode. For now, press CANC and return Pitot to Measure mode. Press ‘4’ to select Pitot mode again. Then press ‘3’ to select Pitot Control mode. The screen appears as shown in Figure 3.17 below. Figure 3.17 : Pitot Control mode ‘Control’ will be blinking in the Pitot mode field. Also, ‘Control’ will be blinking in the Static mode field. This is because when either Pitot mode or Static mode is changed to Control, both Pitot and Static modes are automatically setup for Control. (Caution: Do NOT press GO at this moment). Pressing GO will activate both Pitot Control mode and Static Control mode. Pressing CANC will revert both Pitot and Static modes back to Measure. For now, press CANC and return Pitot and Static to Measure mode. Note: When Pitot mode is changed to Control, Static mode is automatically changed to Control. Similarly, when Static mode is changed to Control, Pitot mode is automatically changed to Control. Note: Leak and Measure modes can be selected independently for Pitot and Static. Static Mode The Static mode can be changed when the key# symbol (8) is displayed to the right of the Static mode field. Press key ‘8’ on the keypad and the Lower section of the display will appear as shown in Figure 3.18 below. 125-9101A Page 37