Download "user manual"

Transcript
2  line/loop impedance
MI 3105 EurotestXA: Measurements
2  line/loop impedance measuring procedure







Connect Impedance adapter to the instrument (see figure 5.49).
Select the functions Z-LINE or Z-LOOP.
Enable and set limit value (optional).
Power ON the Impedance adapter (ON/OFF key, green LED lits).
Connect Impedance adapter to tested installation.
Press the TEST key for measurement.
Store the result (optional).
Figure 5.50: Example of 2  line/loop measurement results
Displayed results:
Z ........................ Line / loop impedance,
ISC ...................... Prospective short-circuit current,
R........................ Resistive part of line impedance,
Xl ....................... Reactive part of line impedance.
The following parameters are displayed in sub-screen for single-phase line impedance
measurement:
IscMaxL-N ......... Maximum prospective short-circuit current.
IscMinL-N .......... Minimum prospective short-circuit current.
IscStd ................ Standard prospective short-circuit current.
When testing phase-to-phase line impedance the following parameters are displayed in
sub-screen:
IscMax3Ph......... Maximum three-phase prospective short-circuit current.
IscMin3Ph.......... Minimum three-phase prospective short-circuit current.
IscMax2Ph......... Maximum two-phase prospective short-circuit current.
IscMin2Ph.......... Minimum two-phase prospective short-circuit current.
IscStd ................ Standard prospective short-circuit current.
The following parameters are displayed in sub-screen for loop impedance measurement:
IscMaxL-Pe ....... Maximum prospective fault current.
IscMinL-Pe ........ Minimum prospective fault current.
IscStd ................ Standard prospective fault current.
Ub...................... Contact voltage at maximum prospective fault current (Contact
voltage is measured against probe S terminal).
76