Download Akta prime

Transcript
A Electrical and communication connections
Appendix A Electrical and
communication connections
1
7
2
8
3
pH-Ground
Drop Sensor Frac Valve
Recorder
RS-232
Rec. On/off
pH-Probe
Conductivity Flow Cell
UV
9
4
Mains
UV-lamp
l
Voltage
100-120 / 220-240 V ~ autorange
Frequency
50-60 Hz
0
l
0
5
10
Power max
90 VA
11
6
(SYSTEM NO.)
(CODE NO.)
No.
Connection
No.
Connection
1
RS-232 to computer
7
pH electrode
2
Flow diversion valve
8
Conductivity flow cell
3
Fraction collector
9
Optical unit
4
Measurement data to recorder
10
Power switch
5
On/off signals to recorder
11
Power inlet
6
UV lamp
Figure A.1: Electrical and communication connections for ÄKTAprime plus.
ÄKTAprime plus Operating Instructions 28-9597-89 AA
57