6.3.3 Delete All Memory Addresses – Data Function
Up to 250 data records can be stored to memory. The memory is full
when the bar graph to the right of “MEMORY:” is entirely filled in. The entire
memory, i.e. all data records from all memory addresses, can be deleted
at once. We recommend uploading and saving your data to a PC before
deletion.
Ð Select Data.
By simultaneously activating O and K all data that have been saved
are deleted from memory. The bar graph to the right of the parameter
„MEMORY:“ is empty. To the left, memory address „001“ is displayed.
New data can now be entered for the first address, or the
database can be exited (press  or START 9 times).
+
3x
Ð
Ð
Enter memory address “000”. After acknowledgement by pressing
button START (press at center) a safety request appears.
If the above message appears upon switching on the test instrument, all
data can be saved to a PC before finally deleting the database in order to
rectify the error.
6.4 Print Function
Functions the symbols of which appear in gray or are displayed only in a
faint raster screen, will not be available until after the next software
update.
+