GasAlert Extreme
User Manual
Clearing Gas Exposures
Table 9. Factory Alarm Setpoints
The peak gas exposures automatically clear after deactivating the detector.
Gas
O2
Low
High
N/A
TWA
N/A
STEL
19.5% vol.
22.5% vol.
35 ppm
200 ppm
35 ppm
200 ppm
35 ppm
200 ppm
35 ppm
200 ppm
10 ppm
15 ppm
10 ppm
15 ppm
To clear the MAX, TWA, and STEL peak exposure readings immediately,
press and hold C for 6 seconds. The detector beeps and vibrates two
times to confirm that the exposures have been cleared.
CO
a Caution
H2S
Follow all safety procedures as defined by your employer. Confirm
with your supervisor before clearing TWA and STEL alarms.
H 2S
10 ppm
15 ppm
10 ppm
15 ppm
PH3
0.3 ppm
1.0 ppm
0.3 ppm
1.0 ppm
Resetting Gas Alarm Setpoints
SO2
2.0 ppm
5.0 ppm
2.0 ppm
5.0 ppm
Cl2
0.5 ppm
1.0 ppm
0.5 ppm
1.0 ppm
NH3
25 ppm
35 ppm
25 ppm
50 ppm
NH3
25 ppm
35 ppm
25 ppm
50 ppm
2.0 ppm
5.0 ppm
2.0 ppm
5.0 ppm
10.0 ppm
Note
Standard factory alarm setpoints vary by region.
Table 9 lists the factory alarm setpoints.
To change the factory alarm setpoints, refer to Calibration and Setting
Alarm Setpoints.
Note
To disable an alarm, set the alarm setpoint to 0.
The ETO sensor is extremely cross sensitive and it responds
strongly to CO.
CO
(low H2)
(high range)
(high range)
NO2
HCN
4.7 ppm
10.0 ppm
4.7 ppm
ETO
1.0 ppm
5.0 ppm
1.0 ppm
5.0 ppm
ClO2
0.1 ppm
0.3 ppm
0.1 ppm
0.3 ppm
O3
0.10 ppm
0.10 ppm
0.10 ppm
0.20 ppm
NO
25 ppm
25 ppm
25 ppm
25 ppm
Sensor Alarm
The detector tests for a missing or defective sensor during the activation
self-test. Refer to Troubleshooting.