APPENDIX E
OPTION 4200-06
E-30. In addition to syntax errors, the
following messages may be generated:
a.
No error:
<so>dddd<cr><If>
Instrument not SETUP: FO7PWMOn (MOD):
b.
POWERMETER NOT SETUP <cr><If>
E-38.
In addition to syntax errors, the
following messages may be generated:
a.
No error: <sp><cr><1f>
Improper channel command:
FO7PWMOn
(DEV): ILLEGAL CHANNEL PROGRAMMED
b.
POWERMETER <cr><If>
Instrument not CLOSED: FO7PWMOn (MOD):
POWERMETER NOT CONNECTED <cr><1f>
c.
c.
Instrument not SETUP:
FO7PWMOn
POWERMETER NOT SETUP <cr><If>
(MOD):
Note: d = ascii numeric digit (0 - 9)
RESULT FETCH. The measurement value
E-31.
is requested by the command:
E-39.
RESET. The RESET command addresses a
specific channel, clears any pending error
mess-ages and sets the 4200 to not-SETUP status. The syntax is:
FTH POWR <cr><If>
RST ACS POWR :CHn<cr><If>
E-32.
The instrument will determine if a
settled reading was possible within the
settling time limit reported to the
controller during the 1NX command. The
measurement value is tested against the maximum and minimum limits established during the
SETUP command. If no errors occur, the
reading is loaded into the output buffer;
otherwise, an error message is loaded.
Execution of the FTH command also
E-33.
returns the 4200 to CLOSED and not-INITIATED
status. Any number of successive INX - FTH
measurement cycles may be performed while
still CLOSED.
In addition to syntax errors, the
E-34.
following messages may be generated:
a.
No error:
<sp>0.ddddEsd<cr><If>
Instrument not SETUP; FO7PWMOn (MOD):
b.
POWERMETER NOT SETUP <cr><If>
Instrument not CLOSED: FO7PWMOn (MOD):
c.
POWERMETER NOT CONNECTED <cr><If>
d. Measurement not INITIATED: FO7PWMOn
(MOD): MEASUREMENT NOT INITIATED <cr><If>
e.
Measurement did not settle: FO5PWMOn
<cr><1f>
(DEV): MEASUREMENT TIMEOUT
f.
E-40.
An illegal channel error message is
generated if the channel addressed does not
exist.
E-41.
The IEEE-488 bus command DCL (Device
Clear) will perform a RESET on the currently
active channel.
E-42.
In addition to syntax errors, the
following messages may be generated:
a.
No error: <so><cr><1f>
Improper channel command:
FO7PWMOn
(DEV): ILLEGAL CHANNEL PROGRAMMED
b.
POWERMETER <cr><If>
E-43. SELF TEST. Two self-test commands are
recognized by the 4200, but the same internal
test is done for both. The test is a checksum verification of the 4200 program memory.
The command syntax is:
CNF<cr><If> or IST<cr><If>
E-44. Each of these commands also does an
implied RESET, returning the 4200 to
not-SETUP status, and clearing any pending
errors prior to self-test.
E-45. In addition to syntax errors, the
following messages may be generated:
Measurement exceeded maximum limit:
FOOPWMOn (DEV): MEASURED POWER GREATER THAN
a.
No error: <sp><cr><If>
b.
Checksum error, CNF command: FO7PWMOn
MAXIMUM <cr><If>
g.
Measurement less than minimum limit:
(DEV): CONFIDENCE TEST FAILURE <cr><If>
FO1PWMOn (DEV): MEASURED POWER LESS THAN
MINIMUM <cr><If>
Checksum error, 1ST command:
FO7PWMOn
(DEV): BIT TEST FAILURE RAM/ROM <cr><If>
c.
Notes: s = exponent sign, +/DISCONNECT. The 4200 may be logically
E-35.
disconnected from the test system by the command:
OPN :CHn<cr><If>
The channel Identification must agree
E-36.
with the currently active channel; otherwise
an illegal channel error will be generated.
The channel must have been SETUP prior to
Issuing this command.
After execution of the OPN command, it
E-37.
is necessary to send the CLS command prior to
INX and FTH commands.
E-46. STATUS COMMAND. Messages in the output buffer must be enabled prior to being
sent to the controller. Except after an INX
or FTH command, which implicitly enable the
output buffer, the STA command must be sent
to access the current message. When the STA
command is received by the 4200, the no-error
message is loaded unless an error message is
already present, and the buffer is enabled.
A subsequent talk command on the IEEE-488 bus
will cause the message to be sent to the
controller.
E- 3