Download BOONTON - emctest.it
Transcript
APPENDIX E OPTION 4200-06 E-30. In addition to syntax errors, the following messages may be generated: a. No error: <so>dddd<cr><If> Instrument not SETUP: FO7PWMOn (MOD): b. POWERMETER NOT SETUP <cr><If> E-38. In addition to syntax errors, the following messages may be generated: a. No error: <sp><cr><1f> Improper channel command: FO7PWMOn (DEV): ILLEGAL CHANNEL PROGRAMMED b. POWERMETER <cr><If> Instrument not CLOSED: FO7PWMOn (MOD): POWERMETER NOT CONNECTED <cr><1f> c. c. Instrument not SETUP: FO7PWMOn POWERMETER NOT SETUP <cr><If> (MOD): Note: d = ascii numeric digit (0 - 9) RESULT FETCH. The measurement value E-31. is requested by the command: E-39. RESET. The RESET command addresses a specific channel, clears any pending error mess-ages and sets the 4200 to not-SETUP status. The syntax is: FTH POWR <cr><If> RST ACS POWR :CHn<cr><If> E-32. The instrument will determine if a settled reading was possible within the settling time limit reported to the controller during the 1NX command. The measurement value is tested against the maximum and minimum limits established during the SETUP command. If no errors occur, the reading is loaded into the output buffer; otherwise, an error message is loaded. Execution of the FTH command also E-33. returns the 4200 to CLOSED and not-INITIATED status. Any number of successive INX - FTH measurement cycles may be performed while still CLOSED. In addition to syntax errors, the E-34. following messages may be generated: a. No error: <sp>0.ddddEsd<cr><If> Instrument not SETUP; FO7PWMOn (MOD): b. POWERMETER NOT SETUP <cr><If> Instrument not CLOSED: FO7PWMOn (MOD): c. POWERMETER NOT CONNECTED <cr><If> d. Measurement not INITIATED: FO7PWMOn (MOD): MEASUREMENT NOT INITIATED <cr><If> e. Measurement did not settle: FO5PWMOn <cr><1f> (DEV): MEASUREMENT TIMEOUT f. E-40. An illegal channel error message is generated if the channel addressed does not exist. E-41. The IEEE-488 bus command DCL (Device Clear) will perform a RESET on the currently active channel. E-42. In addition to syntax errors, the following messages may be generated: a. No error: <so><cr><1f> Improper channel command: FO7PWMOn (DEV): ILLEGAL CHANNEL PROGRAMMED b. POWERMETER <cr><If> E-43. SELF TEST. Two self-test commands are recognized by the 4200, but the same internal test is done for both. The test is a checksum verification of the 4200 program memory. The command syntax is: CNF<cr><If> or IST<cr><If> E-44. Each of these commands also does an implied RESET, returning the 4200 to not-SETUP status, and clearing any pending errors prior to self-test. E-45. In addition to syntax errors, the following messages may be generated: Measurement exceeded maximum limit: FOOPWMOn (DEV): MEASURED POWER GREATER THAN a. No error: <sp><cr><If> b. Checksum error, CNF command: FO7PWMOn MAXIMUM <cr><If> g. Measurement less than minimum limit: (DEV): CONFIDENCE TEST FAILURE <cr><If> FO1PWMOn (DEV): MEASURED POWER LESS THAN MINIMUM <cr><If> Checksum error, 1ST command: FO7PWMOn (DEV): BIT TEST FAILURE RAM/ROM <cr><If> c. Notes: s = exponent sign, +/DISCONNECT. The 4200 may be logically E-35. disconnected from the test system by the command: OPN :CHn<cr><If> The channel Identification must agree E-36. with the currently active channel; otherwise an illegal channel error will be generated. The channel must have been SETUP prior to Issuing this command. After execution of the OPN command, it E-37. is necessary to send the CLS command prior to INX and FTH commands. E-46. STATUS COMMAND. Messages in the output buffer must be enabled prior to being sent to the controller. Except after an INX or FTH command, which implicitly enable the output buffer, the STA command must be sent to access the current message. When the STA command is received by the 4200, the no-error message is loaded unless an error message is already present, and the buffer is enabled. A subsequent talk command on the IEEE-488 bus will cause the message to be sent to the controller. E- 3