Example
Readings 153µm, 150µm, 156µm -> mean value X = 153µm
Standard deviation s = ±3µm
Measuring uncertainty: µ = ± (1 % of reading + 1µm)
D = 153 ±3 ±(1,53µm + 1µm)
= 153 ±5,5µm
LIMITS FUNCTION
Limits can be set in DIRECT and a required GROUP memory at any
time, i.e. before, during and after a series of measurements. There is
practical use for limits:
Any reading which falls outside the tolerance limits set will be registered by a warning indication:
„H“ if reading is above HI limit
„L“ if reading is below LO limit
Limit values can be set via the menu.
GENERAL INFORMATION ON STATISTICS
The instrument calculates statistics from a maximum of 80 readings
per group (GRO1~GRO4 = in total a maximum of 320 readings).
In addition, readings cannot be stored in DIR mode, but it can calculate
statistics as GRO1~GRO4. When the instrument is switched off or
turned to another working mode the DIR values will be lost.
The following statistical values are calculated:
NO = number of readings taken in working mode
AVG = average value
Sdev. = standard deviation
MAX = maximum reading
MIN = minimum reading
41