Download VPF-710- User Manual

Transcript
Product Overview
Section 7
Maintenance
Function
Details
MTBF (Calculated)
52,500 hrs (6 years)
Calibration Check
6 months recommended
Clean Windows
3 months recommended, dependent on
environment
Remote Self-Test Monitoring
Included
VPF750 Specific Temperature and Humidity Measurements
Function
Details
Humidity:
Type
Range
Accuracy
Response Time
Capacitive
0 to 100% Relative Humidity
± 2 % Relative Humidity
< 20 seconds
Temperature:
Type
Range
Accuracy 10°C to 40°C
Accuracy <10°C, >40°C
Response Time
PT 100 Class B
-30°C to 70°C
± 0.2 °C
± 0.2 °C ±0.0073°C/°C
< 20 seconds
Table 7-6 - Sensor Specifications
7.4.1
Instrument Characteristics
Physical
Function
Details
Scattering Angle Coverage
39 to 51
Sample Volume
400 cm3
Weight
5.6Kg
6.8Kg
Length
94
DC Sensors
Mains Powered Sensors
0.75 m
Sensor Specifications