Download PA1000 Power Analyzer
Transcript
Connecting signals To connect a simple current transformer To use a conventional current transformer (CT) like the Tektronix CL series (or any other transducer with a current output), connect the normal AHI and ALO inputs of the PA1000 to the outputs of the current transformer. Follow the manufacturers instructions for the safe use and installation of the transducer. Normally the positive or HI output of the transducer will be marked with the point of an arrow or a + symbol. Connect this terminal to the appropriate AHI input of the PA1000. Figure 12: Current transformer connections Current scaling A current transformer produces an output current that is proportional to the load current being measured. To measure the correct current on the PA1000, use the scaling function of the analyzer to scale, or multiply, the CT output current. For example, the CL200 is a 1000:1 CT. When measuring 100 A, its output is 100 mA. To scale this on the PA1000, a scale factor of 1000 must be entered: Press MENU Select ‘Inputs’ and press Select ‘Scaling’ and press Select ‘Amps’ and press Type the new scale factor (1000) Press Press MENU to return to the measurement display. The PA1000 is now ready to make measurements using a CT. 28 PA1000 Power Analyzer