Download view pdf - AMETEK Programmable Power
Transcript
CAL:VOLT:LEV This command can only be used in the calibration mode. It sets the power supply to a voltage calibration point that is then entered with CAL:VOLT[:DATA]. During calibration, two points must be entered and the low-end point (MIN) must be selected and entered first. Command Syntax Parameters Examples Query Syntax Related Commands CALibrate:VOLTage:LEVel <channel> <CRD> <channel>: 1 – 31, channel number <CRD>: MINimum |MAXimum CAL:VOLT:LEV 1 MIN; CAL:VOLT:LEV 1 MAX (None) CAL:VOLT[:DATA]; CAL:STAT CAL:VOLT:SAV This commands reloads the voltage calibration constants to power supply. These constants are not stored in nonvolatile memory until saved with the *SAV <channel> commands. Command Syntax Parameters Examples Query Syntax Related Commands CALirate:VOLTage:SAVe <channel> <NR1> VOLTage:CALibrate:SAVe <channel> <NR1> <channel>: 1 – 31, channel number <NR1>: 1 – 4, Calibration mode 1 Voltage output 2 Current output 3 Voltage read back 4 Current read back CAL:VOLT:SAV 1 2; VOLT:CAL:SAV 1 4 (None) *SAV CHANNEL SUBSYSTEM The commands in this subsystem only exist in Master – Slave power supply system. They allow users to do the following: x x Query individual channel’s model information. Set or query output shut down time. CHAN:MOD? This query returns the model number of a specified channel. Query Syntax Parameters Returned Parameters Examples Related Commands CHANnel:MODel? <channel> MODel? <channel> <channel>: 1 – 31, channel number <CRD> represents model number CHAN:MOD? 1 MOD? 1 (None) 4-18