8.13.7 SENSe:FRESistance:RESolution
Preparations
for use
Introduction
RESISTOMAT®
Model 2316
DESCRIPTION: Sets the resolution of the measurement display.
SYNTAX:
SENSe:FRESistance:RESolution P1
Meaning of parameter Pn
Safety
instructions
Parameter Meaning
P1
Resolution of the measurement display QUERY FORM:
SENSe:FRESistance:RESolution?
A1
RESPONSE: Value
0.0005 -> Low resolution (2000)
0.00005 -> High resolution (20000)
Controls
Meaning of response An
Operating
instructions in
brief
Response Meaning
A1
Resolution of the measurement display Value
0.0005 -> Low resolution (2000)
0.00005 -> High resolution (20000)
NOTE:
Command not allowed in calibration mode
Command is not allowed when a measurement is running.
RESistance can also be used instead of FRESistance.
Appendix
Maintenance and
customer service
SCPI
commands
Controlling
the instrument
remotely
Operation
8.13.8 SENSe:FRESistance:MODE
Page 100
DESCRIPTION: Selects the measurement mode.
SYNTAX:
SENSe:FRESistance:MODE P1
Meaning of parameter Pn
Parameter Meaning
P1
Measurement mode
Value
SINGle
CONTinuous
ALTernate
CCURve
FASTmeasure
QUERY FORM:
SENSe:FRESistance:MODE?
A1
RESPONSE: Meaning of response An
Response Meaning
A1
Measurement mode
Value
SING
CONT
ALT
CCUR
FAST
->
->
->
->
->
Single shot
Continuous measurement
Alternating measurement
Cooling curve
Fast measurement
->
->
->
->
->
Single shot
Continuous measurement
Alternating measurement
Cooling curve
Fast measurement
NOTE:
Command not allowed in calibration mode.
Command is not allowed when a measurement is running.
RESistance can also be used instead of FRESistance.