Download LEVIFLOW Transit-Time Ultrasonic Flowmeter - Cole
Transcript
TM User Manual for LEVIFLOW LFS Flowmeters www.levitronix.com 3.2 Installation of Converter LFC-6C 3.2.1 Overview and Preparation b a Fixation Bars Converter Termination Box Connector Box Fixation Bar Multi-Converter Array Sensor Connectors c Channel Indication d Channel 6 Pin 1 DIN Rail Mounting Re-Zero Button LOW ADRS Channel Selection High ADRS 4 Digit Display Figure 11: 6-channel controller LFC-6C 1. Check if all parts are delivered as shown in Figure 11. For every configuration (single converter as shown in Figure 11a and multi converter array as in Figure 11b) a termination box (resistance to terminate the RS485) and a connector box (wiring) is needed. 2. For communication with RS485 check the address settings of the potentiometers on the front panel of the converter (see Figure 11d). The total address of the first channel is: “High ADRS” x 10 + “Low ADRS” x 1 TM 3. The LEVIFLOW Configuration Software can be used to debug and configure the flowmeter over the RS485. For communication with RS485 the MODBUS protocol can be requested at Levitronix. PL-4502-00, Rev05, DCO# 11-131 13