12. HAZARDOUS DUTY TYPE INSTRUMENTS
12.2 FM
HAZARDOUS LOCATIONS
(1) Technical Data
Applicable Standard:
FM3600, FM3615, FM3810, ANSI/NEMA 250
Explosion proof for Class I, Division 1,
Groups A, B, C & D.
Dust-ignition proof for Class II/III, Division 1,
Groups E, F & G.
“SEAL ALL CONDUITS WITHIN 18 INCHES”
“WHEN INSTALLED IN DIV. 2, SEALS NOT
REQUIRED”
Conduit
Sealing Fitting
Magnetic Flowmeter
18" (457 mm) Max.
F1202.EPS
Figure 12.2.1 Conduit Wiring
Maximum power supply voltage: 250 VAC/130 VDC
Excitation Circuit: 140 V max
Enclosure: NEMA 4X
Temperature Code: T6
Ambient Temp.: –40 °C to +60 °C (–40 °F to
+140 °F)
(2) Installation
WARNING
• All wiring shall comply with National Electrical
Code ANSI/NFPA 70 and Local Electrical
Code.
• In hazardous locations, wiring to be in conduit
as shown in Figure 12.2.1.
• When installed in Division 2, “SEALS NOT
REQUIRED”
(3) Operation
WARNING
• “OPEN CIRCUIT BEFORE REMOVING COVERS.”
• “SEAL ALL CONDUITS WITHIN 18 INCHES” in
hazardous locations.
• When installed in Division 2, “SEALS NOT
REQUIRED”
• Take care not to generate mechanical spark
when access to the instrument and peripheral
devices in hazardous locations.
(4) Maintenance and Repair
WARNING
The instrument modification or parts replacement
by other than authorized representative of
Yokogawa Electric Corporation is prohibited and
will void the approval of Factory Mutual Research
Corporation.
12-3
IM 01R20C02-01E-E
3rd edition, Nov. 2006